Welcome to Takasago Electric, Inc., a leading manufacturer of valves, pumps and manifolds
Zero-Internal-Volume Valve (Patented)
The "internal-volume" is the space between the diaphragm and the body. When the valve closes, this volume is trapped, often reducing the sensitivity of a test instrument. In the old design, the protruding valve seat is hard and prevents the TeflonTM diaphragm from flattening against the valve body.
In the Zero-Internal-Volume system, a special soft inert elastomer (perfluoroelastomer) is used for the valve seat. When the valve closes, the soft protrusion presses flat, eliminating the internal volume.
Features of Zero-Internal-Volume
Cross-contamination prevention
Faster flushing
Perfect ejection of air-bubbles
Reduction of sample or reagent volumes
The use of soft elastic valve seats also prevents particle induced leakage.
Influence of Internal-Volume on accurate analysis
The below links to the results of an investigation into the influence on analytical accuracy that the Zero-Internal-Volume structure had when incorporated into the concentrator (On-line direct concentrator) for a gas chromatograph.An improvement in analytical accuracy was observed.
Influence of the internal-volume upon accurate analysis
Due to the elastic properties of perfluoroelastomer a slight decrease in flow rate has been observed after long periods of energization.
Models incorporating the Zero-Internal-Volume diaphragm
• STV series (2-way valve only)
• MTV series
• MLV series
Comparative Data Table
Model Type Internal Volume Conventional 87 micro liters MTV-2-M6N Zero-Internal-Volume 34 micro liters
*Passage volumes only.
TrackBack
Solenoid Valves >> Diaphragm Isolated Valves|Pinch Valves|Air Operated Valves|Shape Memory Alloy Valve
Miniature Pumps >> Micro Syringe Pumps|Piezoelectric Micro Pumps|Peristaltic Pumps|Metering Pumps
Manifolds and Chips >> Standard Manifolds|Customized Manifolds|Acrylic Manifold-mountable|Bonded PTFE manifold|Film Chips|Plastic Chips
Fittings2 >> Fittings for PTFE tubing